Velocity Edge DualX Manual de servicio Pagina 169

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 400
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 168
Reference, Installation, and Operations Manual Section 5: Startup and meter configuration
3-9000-743 Rev U August 2014
Current Outputs Page 143
Full scale mass rate:
Series 100 Option Board
Series 100 Plus Option Board
(if installed)
Data points affected:
AO1FullScaleMassRate
AO2FullScaleMassRate
Options:
Enter a mass rate (kg/hr or lbm/hr)
Guidelines
This point is only used when the analog output content is mass
rate.
This point specifies the mass rate associated with the maximum
current (20 mA).
For options and guidelines, see Section 6.2.3 and Daniel
MeterLink online help.
Alarm Action:
Series 100 Option Board
Series 100 Plus Option Board
(if installed)
AO1ActionUponInvalidContent
AO2ActionUponInvalidContent
Options:
High (0)
Low (1)
Hold Last Value (239)
Very Low - 3.5 mA (240)
Very High - 20.5 mA (241)
None (251)
Table 5-15 Data Points for Analog (Current) Output Configuration
Daniel MeterLink
Display Name
Data Points, Options and Guidelines
Vista de pagina 168
1 2 ... 164 165 166 167 168 169 170 171 172 173 174 ... 399 400

Comentarios a estos manuales

Sin comentarios