Reference, Installation, and Operations Manual Section 5: Startup and meter configuration
3-9000-743 Rev U August 2014
Ethernet initial connection steps 105
8. Use the File>Meter Directory menu path to create a new meter directory record with the
following parameters:
Figure 5-1 Program directory
a. Select
Meter Type as Mark III.
b. Check the
Ethernet box and uncheck the Direct and Modem boxes.
c. Click the button and the Ethernet Connections dialog displays. Set
the Ethernet IP address to 192.168.135.100. Click the
OK button to apply the
settings and return to the Meter Directory window
Figure 5-2 Ethernet connections parameters